Abstract

The objective of this study was to investigate the variation of measurements recorded when four different users of the TracerCAD system trace a model of known dimensions and volume. This complements a previous study where the accuracy and consistency of a single user was measured. Landmarks were added to indicate proximal, distal, anterior, medial and lateral regions of a specially manufactured cylindrical nylon 6.6 model. Four circumferential lines were added at regular intervals along the length of the cylinder with a view to calculating diameters and volumes relative to these landmarks. The model was measured using a comparator with guaranteed accuracy to one hundredth of a millimetre, and was traced using the TracerCAD system by four different users. The difference in mean volume between measured results and TracerCAD scans of differing users ranged to -3%. Individual trace volumes varied by up to -7.85%. In all volumes measured, 11 out of 12 maximum volume percentage differences measured greater than 2%, and of these, seven results showed maximum volume percentage difference to measure greater than 4%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.