Abstract

To compare the primary stability of miniscrews after repeated cycles of insertion through insertion torque (IT) measurements and resonance frequency analysis (RFA). Sixty titanium miniscrews were divided into two groups according to the insertion protocol: one with predrilled sites and the other self-drilled into porcine iliac crest bone specimens. Each group had three cycles of reinsertion. After each insertion, the IT and RFA were measured. The IT was measured by using a torque meter, and the RFA was measured using the Osstell ISQ device. A total of five miniscrews of each group were selected for sequential assessment of the morphology of their tip and threads using scanning electron microscopy after each insertion cycle. No statistically significant differences were found in the IT values of miniscrews reinserted up to three times in the group with predrilled surgical sites. The IT value increased significantly with the number of reinsertions in the self-drilled group. The RFA value decreased as the number of insertions increased in both groups. Under the conditions of this in vitro study, reinserting miniscrews deteriorates the integrity of their tip and thread. Reinsertion should be discouraged particularly when insertion sites are not predrilled.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call