Abstract
Technical performances of two optically stimulated luminescence readers including, a microStar reader and an Auto 200 reader both manufactured by Landauer, Inc. were studied by the National Radiation Protection Agency Dosimetry Laboratory using manufacturer and International Electrotechnical Commission requirements. The two reader tools demonstrated satisfactory results as they metall the requirements. The two readers showed similar performance for the overall uncertainty and the Coefficient of Variation tests. The microStar reader showed better performances for the linearity and quality control tests, whereas the Auto 200 reader has a lower limit of detection. Moreover, an average shift of 0.08mSv exists when measuring the doses with the two readers.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.