Abstract

The performance of isocratic separations of 11 pollutant phenols (PP) using monolithic (Chromolith RP-18e) and conventional reversed-phase 5 μm (Luna and Purospher C 18) and 4 μm (Synergi C 12) particulate size columns, selected from high purity silica materials, has been compared. The separations have been optimized based on a previously optimized separation in which a reversed-phase C 18 Luna column and acetonitrile as organic modifier were used, allowing the separation of all phenols tested in 23 min. The optimization process was carried out for each column by studying the effect of the mobile phase (acetonitrile as organic modifier, pH, flow-rate) on phenols separation. Under the optimized separation conditions, all phenols were separated in less than 23 min for all columns tested. Asymmetry factors were further evaluated and used to estimate column efficiency using the Dorsey–Foley equation. The efficiency and asymmetry factors were lower for Chromolith than for Purospher and Luna columns respectively. The Chromolith column was finally selected, due to its lower flow resistance, analysis time and good efficiency and asymmetry factors. The PPs separation was achieved in 3 min. The asymmetry factors were in the range 0.9–1.5 using 50 mM acetate buffer (pH = 5.25)–ACN (64:36, v/v) as mobile phase, T = 45 °C and 4.0 ml min −1 flow-rate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.