Abstract
A phakic intraocular lens (PIOL) of − 4.5 D was characterized from its wavefront aberration profile. A preclinical study was conducted using pre- and post-surgery data from four patients that had undergone myopic laser refractive surgery. All these patients would have needed a PIOL of − 4.5 D. Pre-surgery data were used to simulate the effect of a PIOL implantation. Post myopic refractive surgery data were used to calculate the post-LASIK eye model. Modulation transfer function (MTF), point spread function (PSF) and simulation of optotypes vision were obtained and compared. The PIOL did not worsen the optical quality of the eyes evaluated. High order Aberrations were always higher in the post-LASIK eye model. Optics quality trended to be better in PIOL implantation than post-LASIK surgery as pupil size increased.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.