Abstract

Managing difficult pediatric airway is challenging. The MultiViewScope (MVS) Stylet Scope is reported to be useful in difficult pediatric airway. In this randomized crossover study, we compared the effectiveness of the MVS Stylet Scope to a standard direct laryngoscope with Miller #1 blade in simulated normal and difficult airways. Fifteen expert anesthesiologists and Fifteen anesthesiology residents participated in the study. Participants were asked to perform intubation with the Airsim Baby manikin first, and then with the Airsim Pierre Robin manikin. Participants in each group used the intubation devices in a randomized order. The primary outcome was the time of successful intubation. The secondary outcomes were the force exerted on the incisors during intubation, Cormack-Lehane scale, the difficulty of intubation. There were no differences between MVS Stylet Scope and Direct laryngoscope in the time of successful intubation by the expert anesthesiologists or the anesthesiology residents in a normal or difficult pediatric airway. MVS Stylet Scope significantly improved the force exerted on the incisors during intubation in the expert anesthesiologists or the anesthesiology residents in a normal or difficult pediatric airway. MVS Stylet Scope significantly improved Cormack-Lehane scale, and the difficulty of intubation with difficult pediatric airway situation in both expert anesthesiologists and anesthesiology residents. Although less forces on the incisors and improved view of glottis were observed with the MVS Stylet Scope, MVS Stylet Scope did not shorten the time of intubation. The results of this study mean that the MVS Stylet Scope may be a less invasive airway devise than the direct laryngoscope with the Miller blade in the pediatric airway management. For the next step, we need to evaluate the MVS Stylet Scope in the real patients as an observational study.

Highlights

  • Managing difficult pediatric airway is challenging especially under two years of age, because of the following physiological features [1]

  • MVS Stylet Scope significantly improved the force exerted on the incisors during intubation in the expert anesthesiologists or the anesthesiology residents in a normal or difficult pediatric airway

  • Less forces on the incisors and improved view of glottis were observed with the MVS Stylet Scope, MVS Stylet Scope did not shorten the time of intubation

Read more

Summary

Introduction

Managing difficult pediatric airway is challenging especially under two years of age, because of the following physiological features [1]. Neonates and infants become desaturated more rapidly than adults because of the physiological features. Awake intubation is recommended when managing anticipated difficult airway in adults [3,4,5]. Conducting randomized clinical trials in children with difficult airways is very difficult. The MultiViewScope (MVS) Stylet Scope is reported to be useful in difficult pediatric airway. In this randomized crossover study, we compared the effectiveness of the MVS Stylet Scope to a standard direct laryngoscope with Miller #1 blade in simulated normal and difficult airways

Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.