Abstract
Comparisons of the 1 V Josephson array voltage standard (JAVS) of the D.I. Mendeleyev Institute for Metrology (VNIIM), Russia, with that of the Physikalisch-Technische Bundesanstalt (PTB), Germany, were carried out in St. Petersburg in June 1999. The results of the direct comparison agree within an overall standard uncertainty of less than 1 nV. The results also show that direct comparisons are useful for detecting voltage offsets of a few parts in 10/sup 10/. The reason for a small voltage offset in an SIS array was investigated at zero voltage level.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.