Abstract

A comparison of the 10 V Josephson array voltage standard of the Bureau International des Poids et Mesures (BIPM) was made with that of the D I Mendeleyev Institute for Metrology (VNIIM), Russian Federation, in November 2010. For this exercise, option A of the BIPM.EM-K10.b comparison protocol was applied. This option required the VNIIM to provide a reference voltage for measurement by the BIPM using its Josephson standard with its own measuring device. The BIPM array is kept floating from ground.The final results are in very good agreement with the combined relative standard uncertainty of 2.0 parts in 1010 for the nominal voltage of 10 V.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).