Abstract

A comparison of the 10 V Josephson array voltage standard of the Bureau International des Poids et Mesures (BIPM) was made with that of the Czech Metrology Institute (CMI), Czech Republic, in February 2011. For this exercise, option B of the comparison protocol was applied for the nominal voltage of 10 V. The BIPM provided a reference voltage for measurement by the CMI using its Josephson standard with its own measuring device (option B of the comparison protocol). The final results are in good agreement within the combined relative standard uncertainty of 10.3 parts in 1010 for the nominal voltage of 10 V.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).