Abstract

Objective: Improvements in SARS-CoV-2 diagnosis with easy, rapid and cost-effective approaches are required to control the COVID-19 pandemic. Antigen tests result in 5 to 30 minutes, providing an advantage over polymerase chain reaction (PCR) in duration. We tested the performance of the i-test COVID-19 rapid antigen test to real-time reverse transcriptase PCR in 200 symptomatic COVID-19 suspected patients. The cycle threshold (Ct) values of the patients were found to be between 21.6 and 34.4. The Ct value of 10 patients who tested positive in the PCR test was >30. We found that the sensitivity and specificity of the antigen test were 80.6 % and 93.7 %, respectively, for samples with a Ct value of <30, and overall agreement between antigen and PCR test was 91.6 % for these samples. i-test COVID-19 rapid antigen test can be used for screening in schools, factories, nursing homes, and everywhere where PCR test is unavailable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.