Abstract

Objective: To compare the high-cycle fatigue behavior of four commercially available NiTi orthodontic wires. Material and Methods: Twelve NiTi orthodontic wires, round, 0.016-in, three per brand, were selected and divided into four groups: G1 - Heat-activated NiTi, G2 - Superelastic NiTi, G3 - Therma-Ti, and G4 - CopperNiTi. The atomic absorption spectrometry method was used to determine the chemical composition of investigated NiTi wires. We also performed a fatigue test at three-point bending using a universal testing machine for 1000 cycles in a 35 °C water bath. For the first and thousandth cycle, the average plateau load and the plateau length were determined in the unloading area of the force versus displacement diagram. In addition, we calculated the difference between the average plateau load of the first and thousandth cycle (∆F), as well as the difference between the plateau length of both cases (∆L). Results: According to our results, there were no significant differences between the average plateau load of the first and thousandth cycles of each group (p>0.05) and in the plateau length of the first and thousandth cycles of the groups (p>0.05). Conclusion: There were no significant differences between the groups changing the superelasticity property after high-cycle fatigue.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.