Abstract
Equivalent doses for the eye lenses, thyroid, and mammary glands were measured and compared between one adult and three pediatric anthropomorphic phantoms during chest computed tomography (CT) using 40mm volume helical scan on the Aquilion ONE GENESIS Edition CT equipment. Placing an optically stimulated luminescence dosemeter (OSLD) on the eye lenses, thyroid, and mammary gland, we measured and compared the equivalent dose of OSLD among different phantoms during chest CT using a helical scan. Compared with adults, the equivalent doses to the eye lens, thyroid, and mammary glands were ~81%, 77%, and 63% lower in newborns, 1-year-olds, and 5-year-olds using comparable image noise during chest CT.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.