Abstract

Background/AimsTo prevent the transmission of pathogens by endoscopes, following established reprocessing guidelines is critical. An ideal reprocessing step is simple, fast, and inexpensive. Here, we evaluated and compared the efficacy and safety of two disinfectants, a tertiary amine compound (TAC) and ortho-phthalaldehyde (OPA).MethodsA total of 100 colonoscopes were randomly reprocessed using two same automated endoscope reprocessors, according to disinfectant. The exposure time was 10 minutes for 0.55% OPA (Cidex® OPA, Johnson & Johnson) and 5 minutes for 4% TAC (Sencron2®, Bab Gencel Pharma & Chemical Ind. Co.). Three culture samples were obtained from each colonoscope after reprocessing.ResultsA total of nine samples were positive among the 300 culture samples. The positive culture rate was not statistically different between the two groups (4% for OPA and 2% for TAC, P=0.501). There were no incidents related to safety during the study period.ConclusionsTAC was non-inferior in terms of reprocessing efficacy to OPA and was safe to use. Therefore, TAC seems to be a good alternative disinfectant with a relatively short exposure time and is also less expensive than OPA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.