Abstract
The surface roughness of tooth enamel was quantitatively evaluated following scaling with four power scalers at three different tip forces. Ceramic and tooth samples were evaluated for surface roughness using surface profilometry. Applied tip forces were produced with a horizontal balanced arm holding the scaler handpiece and load weight, and were measured using a load cell. The power scalers evaluated were the ultrasonic piezoelectric, ultrasonic magnetostrictive (ferromagnetic stack), sonic, and rotosonic. For the tooth samples, at 50 grams of force, the piezoelectric, magnetostrictive, and sonic scaler roughnesses were similar and the rotosonic scaler outcome was statistically higher. Increasing the applied force significantly increased the roughness produced by the rotosonic and piezoelectric scalers. The results of this study indicate that the type of power scaler and applied tip force used determines the post-scaling surface roughness of ceramic and tooth enamel surfaces.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.