Abstract

PurposeTo compare the effect of dual-source dual-energy CT versus single-energy CT on DNA double-strand breaks (DSBs) in blood lymphocytes at CT pulmonary angiography (CTPA). Methods and materialsSixty-two patients underwent either dual-energy CTPA (Group 1: n = 21, 80/Sn140 kVp, 89/38 mAs; Group 2: n = 20, 100/Sn140 kVp, 89/76 mAs) or single-energy CTPA (Group 3: n = 21, 120 kVp, 110 mAs). Blood samples were obtained before and 5 min after CTPA. DSBs were assessed with fluorescence microscopy and Kruskal-Walls tests were used to compare DSBs levels among groups. Volume CT dose index (CTDIvol), dose length product (DLP) and organ radiation dose were compared using ANOVA. ResultsThere were increased excess DSB foci per lymphocyte 5 min after CTPA examinations in three groups (Group 1: P = .001; Group 2: P = .001; Group 3: P = .006). There were no differences among groups regarding excess DSB foci/cell and percentage of excess DSBs (Group 1, 23%; Group 2, 24%; Group 3, 20%; P = .932). CTDIvol, DLP and organ radiation dose in Group 1 were the lowest among the groups (all P < .001). ConclusionDSB is increased following dual-source and single-source CTPA, while dual-source dual-energy CT protocols do not increase the estimated radiation dose and also do not result in a higher incidence of DNA DSBs in patients undergoing CTPA.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call