Abstract

The C-MAC comprises a Macintosh blade connected to a video unit. The familiarity of the Macintosh blade, and the ability to use the C-MAC as a direct or indirect laryngoscope, may be advantageous. We wished to compare the C-MAC with Macintosh, Glidescope and Airtraq laryngoscopes in easy and simulated difficult laryngoscopy. Thirty-one experienced anaesthetists performed tracheal intubation in an easy and difficult laryngoscopy scenario. The duration of intubation attempts, success rates, number of intubation attempts and of optimisation manoeuvres, the severity of dental compression, and difficulty of device use were recorded. In easy laryngoscopy, the duration of tracheal intubation attempts were similar with the C-MAC, Macintosh and Airtraq laryngoscopes; the Glidescope performed less well. The C-MAC and Airtraq provided the best glottic views, but the C-MAC was rated as the easiest device to use. In difficult laryngo-scopy the C-MAC demonstrated the shortest tracheal intubation times. The Airtraq provided the best glottic view, with the Macintosh providing the worst view. The C-MAC was the easiest device to use.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.