Abstract

The aim of the present study was to compare bone-collecting capacity of bone harvesting device and minimally irrigated low-speed drilling using three implant systems. One bone harvesting device and three commercially available drill systems were compared using the osteotomies on bovine rib bones. The amount of the collected bone particle and particle size (<500 μm: small, 500–1000 μm: medium, and >1000 μm: large) were measured. Total wet (1.535 ± 0.232 mL) and dry volume (1.147 ± 0.425 mL) of the bone particles from bone harvesting device were significantly greater than three drill systems (wet volume: 1.225 ± 0.187–1.27 ± 0.29 mL and dry volume: 0.688 ± 0.163–0.74 ± 0.311 mL) (P < 0.05). In all groups, the amount of large sized particles in wet and dry state was the greatest compared to that of medium and small particles. The dry weight of the bone particles showed the same tendency to volumetric measurement. In conclusion, total bone particles and large sized particles (>1000 μm) were harvested significantly greater by bone harvesting device than minimally irrigated low-speed drilling. The composition of particle size in all harvesting methods was similar to each other.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.