Abstract
The BluePoint MoldID kit is an oligonucleotide array used for the identification of filamentous fungi, and it has not yet been mentioned in recent studies. We used a BluePoint MoldID kit to identify standard and clinical filamentous fungal isolates and compared its performance with that of Bruker MALDI-TOF MS. The former accurately identified 80.2% of the isolates (142/177), and the latter identified 92.6% of the isolates (164/177). The performance of the BluePoint MoldID kit was slightly inferior to that of Bruker MALDI-TOF MS because of the smaller database. However, the BluePoint MoldID kit can cover most clinically common opportunistic fungal infections; thus, it offers an alternative method for laboratories that lack MALDI-TOF MS equipment, as the device is less expensive.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.