Abstract

The comparison of an automatic vision screening machine using the Landolt rings and the usual Snellen Chart was carried out among 123 second grade and 149 fifth grade students in an elementary school in Jerusalem. The sensitivity of the test for a cut-off point of greater than or equal to 6/12 according to the Snellen test was 41.7% (CI = 16.5-71.4) and the specificity was 86.5% (CI 78.1-92.2) for second graders. For fifth graders the values were 50.0% (CI = 20.1-79.9) and 90.6% (CI 83.7-94.8) respectively. Diagnosis by a specialist decreased the number of 'false negatives' and confirmed the pathological cases. The automatic test was well accepted by the children. Though more time is needed for assessment using the automatic test, its advantage is that no professional staff are required.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.