Abstract

The oxygen permeability of films is relevant for packaging related and technical applications. An increasingly used test method for the measurement of oxygen permeability is the optical test method, because it allows a simple and cost-efficient measurement setup. This method is based on optical chemical sensors. However, not much is known about its validity. Therefore, method validation is necessary which is subject of this study. The optical method is compared with the carrier gas method for a variety of film samples. In the tested permeability range of 0.5–2500 cm3 (STP)·(m2 d bar)−1 both methods deviated less than 20% for zero and 50% relative humidity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.