Abstract
A descriptive correlational study was used to evaluate the correlation/agreement of oral and axillary temperature measurements to patient core temperatures obtained in the OR. Data collectors recorded oral or axillary patient temperature from 752 patients on admission and discharge from the PACU. Results indicated that there was a moderate correlation between each of the current devices and core temperature, but no agreement between core temperature and either device. Recommendations were made to use just one device throughout the organization, or to use the device used on admission throughout the hospitalization.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.