Abstract

Concerns have been raised within the nuclear medicine field about the accuracy of measurements of 89Sr. The 85Sr impurity present in the 89Sr solution has a significant effect on the response of radionuclide calibrators used in hospitals. A comparison was conducted between the National Physical Laboratory (NPL) and the UK hospital physics community. Only 58% of the results were within 5% of the NPL value. Work at NPL has resulted in correction factors for the NPL secondary standard radionuclide calibrator and for Capintec calibrators, which will produce a significant improvement in performance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.