Abstract

The goal of the study was to compare stroke volume (SV) and respiratory stroke volume variation (SVV) measured by pulse-contour analysis and aortic Doppler. These were measured by pulse-contour analysis and thermodilution (PiCCO) and by aortic pulsed wave Doppler with transoesophageal echocardiography in patients undergoing abdominal aortic surgery. Simultaneous measurements were done at different times of surgery. All data were recorded on PiCCOwin software and videotape and analysed off-line by a blinded investigator. A total of 114 measurements were achieved in 20 patients. There was a good correlation and small bias between the PiCCO and the echo-Doppler values of the mean SV [r=0.885; bias=0.2 (8) ml], and between the minimum [r=0.842; bias=1 (9) ml] and maximum SV [r=0.840; bias=2 (10) ml] values. There is a fair correlation between pulse-contour analysis and aortic Doppler for beat-by-beat measurement of SV but not for calculation of SV respiratory ventilation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.