Abstract

The aim of this study was to assess agreement and repeatability of four ultrasound methods for measuring stroke volume (SV) and cardiac output (CO) in cats. Measurement of SV and CO was performed by the Teichholz method, the Simpson's method (SM), the area length method (ALM) and a volumetric flow method across the aorta (Trace method). For each method, the coefficient of variation (CV) was calculated and agreement was determined by Bland-Altman analysis. The CV was acceptable (<20%) for all parameters, except for SV and CO obtained by SM (28.8% and 22.4%, respectively) and ALM (21.6% and 22.6%, respectively). Narrow limits of agreement were observed between both planimetric methods (SM and ALM). The Trace method was the most repeatable, followed by the Teichholz method. Despite excellent inter-method agreement, neither of the planimetric methods produced results with adequate repeatability. As the Teichholz and Trace methods were acceptably repeatable, and probably gave the most representative values, they appear to be the most useful methods for the measurement of SV and CO in cats. Further investigations are needed to compare the echocardiographic methods described here with a standard technique such as thermodilution.