Abstract

The current study, compares the mechanical performance of four modular TKA prostheses based on von Mises stress distribution in the tibial insert. Three-dimensional finite element (FE) models of a cruciate retaining type modular prosthesis and three posterior stabilized (PS) type modular prostheses namely: anterior slide, modular post and double cam, were developed. A compressive load of 2600 N was applied to the FE models at flexion angles of 00, 150, 300, 600 and 900. Von Mises stress was evaluated on all the modular parts of the prostheses and compared with the yield strength of the corresponding material. Von Mises stress in all the parts were below the yield strength of their corresponding material except for tibial insert of anterior slide design at high flexion angle. Von Mises stress above the yield strength in the tibial insert of anterior slide design, was due to edge loading in the post and it demonstrates the likelihood of mechanical failure by delamination type of wear.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.