Abstract

Our aim was to compare the diagnostic performance of strain elastography (SE) and shear-wave elastography (SWE), combined with B-mode ultrasonography (US), in breast cancer. For 79 breast lesions that underwent SE and SWE, two radiologists reviewed five data sets (B-mode US, SWE, SE and two combined sets). Qualitative and quantitative elastographic data and Breast Imaging Reporting and Data System (BI-RADS) categories were recorded. The area under the receiver operating characteristic curve (AUC) was evaluated. No significant difference in the AUC between the two elastography methods was noted. After subjective assessment by reviewers, the AUC for the combined sets was improved (SWE, 0.987; SE, 0.982; B-mode US, 0.970; p < 0.05). When SE and SWE were added, 38% and 56% of benign BI-RADS category 4a lesions with a low suspicion of cancer were downgraded without false-negative results, respectively. SE and SWE performed similarly. Therefore, addition of SE or SWE improved the diagnostic performance of B-mode US, potentially reducing unnecessary biopsies.