Abstract

A supersonic jet spectrum of 9,10-dichloroanthracene is measured by stimulated-emission-pumping fluorescence dip spectrometry and conventional fluorescence spectrometry. The performance obtained is compared for these spectrometric methods, providing same information concerned with the energy level of the ground state. The former is more preferential for measurement of a high-resolution spectrum, since the spectral resolution is determined by the linewidth of the dumping laser. On the other hand, the latter is more preferential for measurement with better sensitivity at the expense of the spectral resolution, since the fluorescence throughput can be improved by increasing the slit width of the monochromator.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.