Abstract
The purpose of this experimental study was to compare the stability of titanium and absorbable plate and screw fixation systems for mandibular angle fractures. Twenty-one sheep hemimandibles were used to evaluate 3 different plating techniques. The groups were fixated with a single titanium plate, a single absorbable plate and double absorbable plates. A cantilever bending biomechanical test model was used for the samples. Each group was tested with vertical forces by a servohydraulic testing unit. The displacement values in each group at each 10 N stage up to 100 N were compared using the 2-way analysis of variance test. The displacement values for the 3 groups differed significantly (P < .05). The variance analyses showed that titanium plate placement had more favorable biomechanical behavior than others. In addition, the 2 absorbable plates group had more favorable biomechanical behavior than a single absorbable plate group but it was not significantly different at 10 to 40 N. The study demonstrated that titanium plate and screw fixation system had greater resistance to occlusal loads than absorbable plate and screw systems. In addition, a second absorbable plate orientation provides a more favorable biomechanical behavior than a single absorbable plate placement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.