Abstract

Human scalp hair and some kinds of vegetable and animal fibers were analyzed by means of the SR excited X-ray fluorescence method (SRXFA) and the neutron activation method (NAA). Human hair samples collected from five males and five females were washed by the IAEA method prior to analysis. In the SRXFA analysis, samples were excited by monochromated X-rays. Fluorescence X-rays were measured by an Si(Li) detector. The elements detected in all hair samples were S, Ca, Cu, Fe, Zn, Br, and Sr. The elements K, Ti, Cr, Mn, Ni, Se, Hg, and Pb were also detected in several samples. After SRXFA analysis these same samples were analyzed by the NAA method. Elements such as Cu, Zn, and Br were detected by both methods, and their relative concentrations show a good agreement of variation between individuals. However, Pb was only detected by SRXFA, and Na, Au, and Sb were only detected by NAA. Therefore, these two methods are complementary to each other for trace element analysis.