Abstract
This research investigated whether the Sprotte needle causes less leakage of CSF than the Quincke needle in the artificial spinal cord. The changes in intradural pressure, extradural pressure, and leaked volume of CSF were evaluated following puncture with Sprotte and Quincke needles in the artificial spinal cord. The decrease in intradural pressure was 9.7±1.8 mm H2O with the Sprotte needle and 20.5±2.7 mm H2O with the Quincke needle (P<0.05). The volume of leakage of artificial CSF was 2.0±0.3 ml with the Sprotte needle and 3.3 ±0.3 ml with the Quincke needle (P<0.01). The extradural pressure increase was 166.1±8.2 mm H2O with the Sprotte needle and 186.8±13.2 mm H2O with the Quincke needle (P<0.05). The Sprotte needle produces less CSF leakage than the Quincke needle.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.