Abstract
The aim of this study was to compare the effect of the sirolimus loaded PLGA-PEG co-polymer coronary stent with that of the bare metal stent in a porcine coronary restenosis model. Pigs were randomized into two groups (10 pigs, 10 coronary arteries in each group), in which either the sirolimus loaded PLGA-PEG copolymer stent (SPPS, n=10) or the bare metal stent (BMS, n=10) was implanted in the coronary arteries. Histopathologic analysis was performed at 28 days after stenting. There were no significant differences in the injury score, internal elastic lamina (IEL) area, and inflammation score between the two groups. There were significant differences in the lumen area (3.6±0.42 mm2 in the SPPS group vs. 2.5±0.89 mm2 in the BMS group, p<0.001), in the neointimal area (1.2±0.34 mm2 in the SPPS group vs. 2.5±0.99 mm2 in the BMS group, p<0.001), in the percent area stenosis (26.0±6.58% in the SPPS group vs. 48.6±16.95% in the BMS group, p<0.001) and in the fibrin score (1.2±0.41 in the SPPS group vs. 0.5±0.26 in the BMS group, p<0.001). These results demonstrate that SPPS had a superior effect on suppression of neointimal hyperplasia than BMS at one month after stenting in a porcine coronary restenosis model.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.