Abstract
A system of differential equations modeling a heat flux DSC is solved and the results are compared with those obtained using a TA Instruments Q1000™ DSC. 1 US Patent 6,431,747 additional patents applied for. 1 It incorporates a new heat flow rate measurement technique that determines the heat flow rate between the sample and its pan. Two types of first-order transitions are investigated: melting of a pure substance and solidification of a pure substance including super-cooling. In both transitions, the peak shape obtained using the new heat flow rate measurement and predicted by the model is quite different from that measured using conventional DSC. It is shown that the differences are the result of simplifications implicit in the conventional heat flow rate measurement that is based solely on the difference between sample and reference calorimeter temperatures. Heat flow rates measured using the improved measurement agree very well with the model predictions for heat exchange between the sample and its pan.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.