Comparison of RT-PCR, RT-LAMP, and Antigen Quantification Assays for the Detection of SARS-CoV-2
A rapid and simple alternative test to real-time reverse transcription-polymerase chain reaction (RT-PCR) is required for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) to help curb the spread of coronavirus disease (COVID-19). In the present study, we compared the RT-PCR method with chemiluminescent enzyme immunoassay (CLEIA) and reverse transcription loop-mediated isothermal amplification (RT-LAMP). We observed that the number of SARS-CoV-2 RNA copies and the CLEIA antigen quantification values were highly correlated. The detection limit for antigen quantification was 42.8 RNA copies for saliva samples and 23.4 copies for nasopharyngeal swab samples. For both purified RNA and purification-free crude RNA, the number of RNA copies and RT-LAMP threshold time (Tt) values were inversely correlated. RT-LAMP with purified RNA detected low copy numbers of RNA (5-50 copies), whereas fewer than 250 RNA copies could not be detected using crude RNA. CLEIA antigen quantification is potentially useful for large-scale screening, as it is compatible with high-throughput testing. RT-LAMP with crude RNA samples is applicable for rapid point-of-care testing because it can directly use patient specimens. It is important to select a diagnostic method that is simple and rapid when compared with RT-PCR, depending on the situation.
- # Reverse Transcription Loop-mediated Isothermal Amplification
- # Severe Acute Respiratory Syndrome Coronavirus 2
- # Real-time Reverse Transcription-polymerase Chain Reaction
- # Crude RNA
- # Antigen Quantification
- # Copy Numbers Of RNA
- # RNA Copies
- # Chemiluminescent Enzyme Immunoassay
- # Reverse Transcription Loop-mediated Isothermal Amplification Assays
- # Real-time Reverse Transcription-polymerase Chain Reaction Assays
- Research Article
6
- 10.1186/s12879-023-08924-3
- Jan 15, 2024
- BMC Infectious Diseases
BackgroundThe diagnostic assay leveraging multiple reverse transcription loop-mediated isothermal amplification (RT-LAMP) could meet the requirements for rapid nucleic acid detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2).MethodsThe devised assay merged the lateral flow assay with the RT-LAMP technology and designed specific primers for the simultaneous detection of the target and human-derived internal reference genes within a single reaction. An inquiry into the assay's limit of detection (LOD), sensitivity, and specificity was carried out. The effectiveness of this assay was validated using 498 clinical specimens.ResultsThis LOD of the assay was determined to be 500 copies/mL, and there was no observed cross-reaction with other respiratory pathogens. The detection results derived from clinical specimens showed substantial concordance with those from real-time reverse transcription-polymerase chain reaction (RT-qPCR) (Cohen's kappa, 0.876; 95% CI: 0.833-0.919; p<0.005). The diagnostic sensitivity and specificity were 87.1% and 100%, respectively.ConclusionThe RT-LAMP assay, paired with a straightforward and disposable lateral immunochromatographic strip, achieves visual detection of dual targets for SARS-CoV-2 immediatly. Moreover, the entire procedure abstains from nucleic acids extraction. The samples are lysed at room temperature and subsequently proceed directly to the RT-LAMP reaction, which can be executed within 30 minutes at a constant temperature of 60-65°C. Then, the RT-LAMP amplification products are visualized using colloidal gold test strips.Trial registrationThis study was registered at the Chinese Clinical Trial Registry (Registration number: ChiCTR2200060495, Date of registration 2022-06-03).
- Abstract
9
- 10.1016/s2214-109x(21)00111-x
- Mar 1, 2021
- The Lancet. Global Health
Reverse transcription loop-mediated isothermal amplification (RT-LAMP) for point-of-care detection of SARS-CoV-2: a clinical study to evaluate agreement with RT-qPCR
- Research Article
5
- 10.1016/j.jcvp.2023.100134
- Jan 27, 2023
- Journal of Clinical Virology Plus
Reverse transcription-loop-mediated isothermal amplification (RT-LAMP) assay as a rapid molecular diagnostic tool for COVID-19 in healthcare workers
- Research Article
24
- 10.1016/j.jiac.2021.05.004
- May 11, 2021
- Journal of Infection and Chemotherapy
Evaluation of SARS-CoV-2 RNA quantification by RT-LAMP compared to RT-qPCR
- Research Article
134
- 10.1186/s12985-020-01435-6
- Oct 21, 2020
- Virology Journal
BackgroundFast, reliable and easy to handle methods are required to facilitate urgently needed point-of-care testing (POCT) in the current coronavirus pandemic. Life-threatening severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has rapidly spread all over the world, infecting more than 33,500,000 people and killing over 1 million of them as of October 2020. Infected individuals without any symptoms might still transfer the virus to others underlining the extraordinary transmissibility of this new coronavirus. In order to identify early infections effectively, treat patients on time and control disease spreading, rapid, accurate and onsite testing methods are urgently required.ResultsHere we report the development of a loop-mediated isothermal amplification (LAMP) based method to detect SARS-CoV-2 genes ORF8 and N directly from pharyngeal swab samples. The established reverse transcription LAMP (RT-LAMP) assay detects SARS-CoV-2 directly from pharyngeal swab samples without previous time-consuming and laborious RNA extraction. The assay is sensitive and highly specific for SARS-CoV-2 detection, showing no cross reactivity when tested on 20 other respiratory pathogens. The assay is 12 times faster and 10 times cheaper than routine reverse transcription real-time polymerase chain reaction, depending on the assay used.ConclusionThe fast and easy to handle RT-LAMP assay amplifying specifically the genomic regions ORF8 and N of SARS-CoV-2 is ideally suited for POCT at e.g. railway stations, airports or hospitals. Given the current pandemic situation, rapid, cost efficient and onsite methods like the here presented RT-LAMP assay are urgently needed to contain the viral spread.
- Research Article
37
- 10.1371/journal.pntd.0008855
- Nov 4, 2020
- PLOS Neglected Tropical Diseases
Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) is a highly pathogenic novel coronavirus that has caused a worldwide outbreak. Here we describe a reverse transcription loop-mediated isothermal amplification (RT-LAMP) assay that uses a portable device for efficient detection of SARS-CoV-2. This RT-LAMP assay specifically detected SARS-CoV-2 without cross-reacting with the most closely related human coronavirus, SARS-CoV. Clinical evaluation of nasal swab samples from suspected SARS-CoV-2 pneumonia (COVID-19) patients showed that the assay could detect over 23.7 copies within 15 min with a 100% probability. Since the RT-LAMP assay can be performed with a portable battery-supported device, it is a rapid, simple, and sensitive diagnostic assay for COVID-19 that can be available at point-of-care. We also developed the RT-LAMP assay without the RNA extraction step–Direct RT-LAMP, which could detect more than 1.43 x 103 copies within 15 min with a 100% probability in clinical evaluation test. Although the Direct RT-LAMP assay was less sensitive than the standard RT-LAMP, the Direct RT-LAMP assay can be available as the rapid first screening of COVID-19 in poorly equipped areas, such as rural areas in developing countries.
- Research Article
72
- 10.1016/j.jcv.2020.104616
- Aug 31, 2020
- Journal of Clinical Virology
Use of the variplex™ SARS-CoV-2 RT-LAMP as a rapid molecular assay to complement RT-PCR for COVID-19 diagnosis.
- Research Article
14
- 10.7171/jbt.21-32-03-009
- Sep 1, 2021
- Journal of Biomolecular Techniques : JBT
Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) control in the United States remains hampered, in part, by testing limitations. We evaluated a simple, outdoor, mobile, colorimetric reverse-transcription loop-mediated isothermal amplification (RT-LAMP) assay workflow where self-collected saliva is tested for SARS-CoV-2 RNA. From July 16, 2020, to November 19, 2020, surveillance samples (n = 4704) were collected from volunteers and tested for SARS-CoV-2 at 5 sites. Twenty-one samples tested positive for SARS-CoV-2 by RT-LAMP; 12 were confirmed positive by subsequent quantitative reverse-transcription polymerase chain reaction (qRT-PCR) testing, whereas 8 tested negative for SARS-CoV-2 RNA, and 1 could not be confirmed because the donor did not consent to further molecular testing. We estimated the false-negative rate of the RT-LAMP assay only from July 16, 2020, to September 17, 2020 by pooling residual heat-inactivated saliva that was unambiguously negative by RT-LAMP into groups of 6 or fewer and testing for SARS-CoV-2 RNA by qRT-PCR. We observed a 98.8% concordance between the RT-LAMP and qRT-PCR assays, with only 5 of 421 RT-LAMP-negative pools (2493 total samples) testing positive in the more-sensitive qRT-PCR assay. Overall, we demonstrate a rapid testing method that can be implemented outside the traditional laboratory setting by individuals with basic molecular biology skills and that can effectively identify asymptomatic individuals who would not typically meet the criteria for symptom-based testing modalities.
- Research Article
39
- 10.3389/fmicb.2021.713713
- Nov 18, 2021
- Frontiers in Microbiology
The coronavirus disease 2019 (COVID-19) pandemic unfolded due to the widespread severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) transmission reinforced the urgent need for affordable molecular diagnostic alternative methods for massive testing screening. We present the clinical validation of a pH-dependent colorimetric reverse transcription loop-mediated isothermal amplification (RT-LAMP) for SARS-CoV-2 detection. The method revealed a limit of detection of 19.3 ± 2.7 viral genomic copies/μL when using RNA extracted samples obtained from nasopharyngeal swabs collected in guanidine-containing viral transport medium. Typical RT-LAMP reactions were performed at 65°C for 30 min. When compared to reverse transcriptase–quantitative polymerase chain reaction (RT-qPCR), up to cycle-threshold (Ct) value 32, RT-LAMP presented 98% [95% confidence interval (CI) = 95.3–99.5%] sensitivity and 100% (95% CI = 94.5–100%) specificity for SARS-CoV-2 RNA detection targeting E and N genes. No cross-reactivity was detected when testing other non–SARS-CoV virus, confirming high specificity. The test is compatible with primary RNA extraction–free samples. We also demonstrated that colorimetric RT-LAMP can detect SARS-CoV-2 variants of concern and variants of interest, such as variants occurring in Brazil named gamma (P.1), zeta (P.2), delta (B.1.617.2), B.1.1.374, and B.1.1.371. The method meets point-of-care requirements and can be deployed in the field for high-throughput COVID-19 testing campaigns, especially in countries where COVID-19 testing efforts are far from ideal to tackle the pandemics. Although RT-qPCR is considered the gold standard for SARS-CoV-2 RNA detection, it requires expensive equipment, infrastructure, and highly trained personnel. In contrast, RT-LAMP emerges as an affordable, inexpensive, and simple alternative for SARS-CoV-2 molecular detection that can be applied to massive COVID-19 testing campaigns and save lives.
- Research Article
537
- 10.1016/j.cmi.2020.04.001
- Apr 8, 2020
- Clinical microbiology and infection : the official publication of the European Society of Clinical Microbiology and Infectious Diseases
Rapid and visual detection of 2019 novel coronavirus (SARS-CoV-2) by a reverse transcription loop-mediated isothermal amplification assay
- Research Article
13
- 10.1016/j.aca.2022.340036
- Jun 6, 2022
- Analytica Chimica Acta
Electromagnetically-driven integrated microfluidic platform using reverse transcription loop-mediated isothermal amplification for detection of severe acute respiratory syndrome coronavirus 2
- Research Article
7
- 10.1128/spectrum.01962-22
- Nov 29, 2022
- Microbiology Spectrum
Accurate and simple diagnostic tests for coronavirus disease 2019 (COVID-19) are essential components of the pandemic response. In this study, we evaluated a one-tube reverse transcription-loop-mediated isothermal amplification (RT-LAMP) assay coupled with clustered regularly interspaced short palindromic repeat (CRISPR)-associated protein-mediated endpoint detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) RNA in clinical samples. RT-LAMP-CRISPR is fast and affordable, does not require bulky thermocyclers, and minimizes carryover contamination risk. Results can be read either visually or with a fluorometer. RT-LAMP-CRISPR assays using primers targeting a highly expressed nsp8 gene and previously described nucleocapsid (N) gene primers were designed. The analytical characteristics and diagnostic performance of RT-LAMP-CRISPR assays were compared to those of a commercial real-time RT-PCR E gene assay. The limits of detection (LODs) of the nsp8 and N RT-LAMP-CRISPR assays were 750 and 2,000 copies/mL, which were higher than that of the commercial real-time RT-PCR assay (31.3 copies/mL). Despite the higher LOD, RT-LAMP-CRISPR assays showed diagnostic sensitivity and specificity of 98.6% and 100%, respectively, equivalent to those of the real-time RT-PCR assay (P = 0.5). The median fluorescence reading from the nsp8 assay (378.3 raw fluorescence unit [RFU] [range, 215.6 to 592.6]) was significantly higher than that of the N gene assay (342.0 RFU [range, 143.0 to 576.6]) (P < 0.0001). In conclusion, we demonstrate that RT-LAMP-CRISPR assays using primers rationally designed from highly expressed gene targets are highly sensitive, specific, and easy to perform. Such assays are a valuable asset in resource-limited settings. IMPORTANCE Accurate tests for the diagnosis of SARS-CoV-2, the virus causing coronavirus disease 2019 (COVID-19), are important for timely treatment and infection control decisions. Conventional tests such as real-time reverse transcription-PCR (RT-PCR) require specialized equipment and are expensive. On the other hand, rapid antigen tests suffer from a lack of sensitivity. In this study, we describe a novel assay format for the diagnosis of COVID-19 that is based on principles of loop-mediated isothermal amplification (LAMP) and clustered regularly interspaced short palindromic repeat (CRISPR)-Cas chemistry. A major advantage of this assay format is that it does not require expensive equipment to perform, and results can be read visually. This method proved to be fast, easy to perform, and inexpensive. The test compared well against an RT-PCR assay in terms of the ability to detect SARS-CoV-2 RNA in clinical samples. No false-positive test results were observed. The new assay format is ideal for SARS-CoV-2 diagnosis in resource-limited settings.
- Research Article
32
- 10.3389/fcimb.2021.653616
- Jun 29, 2021
- Frontiers in Cellular and Infection Microbiology
Coronavirus disease 2019 (COVID-19) caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has become a major threat to public health. Rapid molecular testing for convenient and timely diagnosis of SARS-CoV-2 infections represents a challenge that could help to control the current pandemic and prevent future outbreaks. We aimed to develop and validate a multiplex and colorimetric reverse transcription loop-mediated isothermal amplification (RT-LAMP) assay using lyophilized LAMP reagents for sensitive and rapid detection of SARS-CoV-2. LAMP primers were designed for a set of gene targets identified by a genome-wide comparison of viruses. Primer sets that showed optimal features were combined into a multiplex RT-LAMP assay. Analytical validation included assessment of the limit of detection (LoD), intra- and inter-assay precision, and cross-reaction with other respiratory pathogens. Clinical performance compared to that of real-time reverse transcriptase-polymerase chain reaction (RT-qPCR) was assessed using 278 clinical RNA samples isolated from swabs collected from individuals tested for COVID-19. The RT-LAMP assay targeting the RNA-dependent RNA polymerase (RdRp), membrane (M), and ORF1ab genes achieved a comparable LoD (0.65 PFU/mL, CT=34.12) to RT-qPCR and was 10-fold more sensitive than RT-qPCR at detecting viral RNA in clinical samples. Cross-reactivity to other respiratory pathogens was not observed. The multiplex RT-LAMP assay demonstrated a strong robustness and acceptable intra- and inter-assay precision (mean coefficient of variation, 4.75% and 8.30%). Diagnostic sensitivity and specificity values were 100.0% (95% CI: 97.4–100.0%) and 98.6% (95% CI: 94.9–99.8%), respectively, showing high consistency (Cohen’s kappa, 0.986; 95% CI: 0.966–1.000; p<0.0001) compared to RT-qPCR. The novel one-step multiplex RT-LAMP assay is storable at room temperature and showed similar diagnostic accuracy to conventional RT-qPCR, while being faster (<45 min), simpler, and cheaper. The new assay could allow reliable and early diagnosis of SARS-CoV-2 infections in primary health care. It may aid large-scale testing in resource-limited settings, especially if it is integrated into a point-of-care diagnostic device.
- Research Article
3
- 10.3892/etm.2023.12097
- Jul 6, 2023
- Experimental and Therapeutic Medicine
The detection of coronavirus disease 2019 cases represents a significant challenge at the epidemiological level. Limitations exist in effectively detecting asymptomatic cases, achieving good follow-up in hospitals without the infrastructure for reverse transcription-quantitative PCR (RT-qPCR) or in difficult-to-access areas and developing methods with the need for less invasive sampling procedures. Therefore, the present study evaluated the performance of the direct reverse transcription loop-mediated isothermal amplification (RT-LAMP) test for detecting severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in the saliva and nasal samples of asymptomatic individuals belonging to the university population. In addition, this test was also assessed for effectiveness in symptomatic individuals referred from a hospital with poor infrastructure in molecular biology and located outside the urban area. The RT-LAMP assay was compared with the results obtained from the RT-qPCR nasopharyngeal swab test, where the diagnosis was confirmed by lateral flow immunoassay test for rapid antigen detection. A total of 128 samples were analyzed, of which 43% were symptomatic positive individuals, 25% were asymptomatic positive individuals and 32% were SARS-CoV2-negative control individuals. Among positive individuals, no differences were found between the Cq values determined by RT-qPCR. A sensitivity of 96.5% and a specificity of 97.6% was reported for the detection of SARS-CoV-2 in symptomatic individuals by salivary and nasal RT-LAMP, as well as a sensitivity of 100% and a specificity of 97.6% for the detection of SARS-CoV-2 in asymptomatic individuals. These findings indicated that performance of the direct RT-LAMP test using saliva and nasal samples has high sensitivity and specificity, which in turn suggest that it is a viable and reliable alternative for use in epidemiological monitoring.
- Research Article
11
- 10.1111/ajt.16464
- Feb 18, 2021
- American Journal of Transplantation
Arterial abnormalities identified in kidneys transplanted into children during the COVID-19 pandemic.