Abstract

To compare basal retinal and cerebral blood flow (BF) values using continuous arterial spin labeling (CASL) MRI and fluorescent microspheres. A total of 41 animals were used. BF was measured using an established microsphere technique (a mixture of 2.5 million 8 μm green and 0.5 million 10 μm blue fluorescent microspheres) and CASL MRI blood flow measurement in the rat retina and brain at 7 Tesla (T) and 11.7T, respectively. Retinal BF by MRI was 1.18 ± 0.57 mL/g/min and choroidal BF was 8.14 ± 1.8 mL/g/min (n = 6). Microsphere retinal BF was 9.12 ± 2.8 μL/min per tissue and choroidal BF was 73.38 ± 44 μL/min per tissue (n = 18), corresponding to a retinal BF value of 1.22 ± 0.36 mL/g/min by means of a wet weight conversion. The wet-weight of the choroid could not be determined. To corroborate our findings, cerebral BF (CBF) by MRI was also analyzed. In the cerebral cortices, CBF was 0.91 ± 0.29 mL/g/min (n = 14) by CASL MRI and 1.09 ± 0.37 mL/g/min (n = 6) by microspheres. There were no significant differences found between MRI and microsphere blood flow in the retina and brain. BF values in the rat retina and cerebral cortex by MRI are in agreement with those obtained by the microsphere technique.