Abstract

To compare two swept-source OCT optical biometers, the Anterion® (Heidelberg Engineering GmbH, Heidelberg, Germany) and the IOLMaster 700® (Carl Zeiss Meditec AG, Jena, Germany), in the analysis of biometric data, intraocular lens (IOL) calculation and postoperative spherical equivalent predictability. This was a real-life, single-center, retrospective study including 152 eyes of 81 consecutive patients referred for cataract surgery. All patients were examined with the IOLMaster 700® and the Anterion®. Biometric data (axial length [AL], anterior chamber depth [ACD], mean keratometry [Km], K1 [flat keratometry], K2 [steep keratometry] and axis, TK1 [flat total keratometry], TK2 [steep total keratometry], central pachymetry, lens thickness [LT], white-to-white distance [WTW]), IOL calculation with the SRK/T formula, and postoperative refractive outcome at 1 month were compared. All biometric measurements were significantly different between the two biometers. Correlations were excellent for AL, pachymetry, ACD, LT and keratometry measurements, and for the IOL calculation (r>0.96, intraclass correlation coefficient=1). The IOL power for emmetropia was similar between both biometers when the SRK/T formula was used (20.84±3.24D versus 20.86±3.29D, P=0.61). The mean postoperative spherical equivalent prediction error calculated using the SRK/T formula was 0.03±0.5D for the IOLMaster 700® versus 0.01±0.47D for the Anterion® (P=0.12). This study showed excellent correlation and agreement for the biometric measurements and the IOL power calculation with the SRK/T formula between both biometers.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call