Abstract

OBJECTIVETo compare the sensitivity and specificity of luciferase immunoprecipitation (LIPS) with radioimmunoprecipitation (RIP) for the measurement of autoantibodies to the type 1 diabetes autoantigens glutamic acid decarboxylase 65 (GAD65) and insulinoma-associated protein (IA)-2β.RESEARCH DESIGN AND METHODSSera from 49 type 1 diabetic patients and 100 nondiabetic control subjects from Diabetes Antibody Standardization Program 2007 were used to screen for autoantibodies to GAD65. An additional 200 type 1 diabetic patients and 200 nondiabetic control subjects were used to validate the GAD65 results and screen for autoantibodies to IA-2β.RESULTSLIPS showed equal sensitivity and specificity to RIP for detecting autoantibodies to GAD65 and IA-2β. Receiver-operating characteristic analysis revealed that the detection of autoantibodies to GAD65 and IA-2β by LIPS and RIP were not statistically different.CONCLUSIONSThe LIPS assay does not require the use of radioisotopes or in vitro transcription/translation and is a practical alternative at the clinical level for the RIP assay.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.