Abstract

The purpose of this study was to compare the ability of radiographic and electronic methods to determine reliably the location of the apical constriction. The working length of root canals in human cadaver teeth was determined by positioning an endodontic file at the apical termination point as indicated by an apex locator. These same teeth were radiographed at various angles with the file in place. The radiographs were evaluated by five examiners to determine a radiographic termination point for each canal. Teeth were extracted and examined histologically to determine the ideal termination point. The deviations of the two experimental termination points from the ideal termination point were compared. The mean of the absolute value of the deviations from the apical constriction for the apex locator was significantly less (p < 0.05) than that for the radiographic method. Thus, the method using the apex locator was slightly more reliable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.