Abstract

Objective To evaluate the effective organ radiation dose in patients with conventional coronary angiography(CCA) and XperSwing coronary angiography.Methods The operating environment of CCA and XperSwing coronary angiography was simulated using the Chinese anthropomorphic chest phantom.The thermoluminescent dosimeters (TLDs) were placed on major organs of the phantomm.The organs' absorbed doses were measured and the dose-area products (DAPs) were recorded.The real effective dose was the summation of products of every organs'absorbed dose × a factor.The calculated effective doses was the product of the DAP × a factor.Results The average absorbed effective doses of XperSwing coronary angiography were lower than that of the CCA.The calculated effective doses of CCA and XperSwing coronary angiography were 3.88 mSv and 3.13 mSv,respectively,and less than the real ones (5.57 mSv,4.94 mSv).Conclusions XperSwing coronary angiography would have less radiation dose than CCA does.The calculated effective dose would be less than the real one. Key words: Interventional radiology; Coronary angiography; Radiation dose; Phantom

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.