Abstract
Polyoxyethyleneglycerol triricinoleate 35 (Cremophor EL: CrEL) is a solubilization agent for hydrophobic drugs. Recently, CrEL has shown some side effects in patients. In the present work, we introduce pyrolysis–mass spectrometry (Py–MS) for the determination of CrEL in drugs and blood samples. Mass to charge ( m/ z) values of 89 and 138 of CrEL and 3-nitroaniline (as internal standard), respectively, were used for quantitative measurements by selected ion monitoring (SIM) method. At a probe pyrolysis temperature range of 350–450 °C the results are highly reproducible. Limit of detection (LOD), linearity and relative standard deviation (R.S.D., n=5) were determined to be 1 ng ml −1, 10 ng ml −1–100 mg ml −1 and 1.3%, respectively. The results of Py–MS are compared with those obtained by high performance liquid chromatography (HPLC) and show that time for analysis, sensitivity and linearity are far better.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.