Abstract

The relative merits of eight procedures for preconcentrations of trace metal ions from natural water samples and synthetic solutions are evaluated. Spikes (100 μg l −1 ) of Mn, Co, Zn, Eu, Cs and Ba and the corresponding radioactive tracers were added to batches of drinking water, estuarine water, sea water, ground water, twice-distilled water and ahumic material solution. After equilibration for 2–5 months, the following techniques were applied: passage through columns of Dowex Al chelating resin and ofsilylated silica gel, filtration through laminate membrane filters and chelating diethylenetriamine cellulose filters, precipitation with sodium diethyldithiocarbamate and l-(2-pyridylazo)-2-naphthol, extraction with ammonium pyrrolidinedithiocarbamate, and chelation by 8-quinolinol (oxine) followed by adsorption on activated carbon. The quantitative characteristics of these techniques and the influence of the water matrix effects are discussed, as well as the applicability for x-ray fluorescence analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.