Abstract

We have evaluated the potential of monoclonal antibodies in the development of a non-isotopic immunometric assay for intact human parathyroid hormone (PTH). The assay has been designed to utilise a chemiluminescent acridinium ester labelled anti-aminoterminal (anti-N) antibody and a solid-phase anti-carboxyterminal antibody in order to measure specifically the intact hormone. In this system the characteristics of the labelled antibody proved crucial to the performance of the assay. A low affinity monoclonal reagent yielded insufficient analytical sensitivity, while a higher affinity monoclonal reagent cross-reacted poorly with the intact molecule relative to the amino terminal PTH fragment to which it was raised. Neither antibody could therefore match the performance of an affinity-purified polyclonal anti-N PTH reagent. These results highlight the problems to be addressed in the selection of suitable reagents for immunometric assay development when specificity and sensitivity are crucial requirements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.