Abstract

Homozygous or severe heterozygous familial hypercholesterolemia and elevated lipoprotein(a) levels may be treated with membrane filtration. The MONET system (Fresenius Medical Care, Bad Homburg, Germany) involves plasma separation by centrifugation or filtration. Whether the method of plasma separation affects lipoprotein lowering and treatment safety was investigated in a single-center retrospective study. The centrifugation-based plasma separation achieved a higher plasma flow and shorter time to treat 1L of plasma (46.2 ± 8.6min), than the filtration-based system (71.5 ± 40.0min; p=0.001). The mean reduction of LDL-cholesterol was 69% and 67% with centrifugation and filtration and was 75% for lipoprotein(a) with both plasma separation methods. A reduction of IgM by more than 60%, of albumin and total protein by approximately 20% and low frequency of side effects was observed. The efficacy of lowering atherogenic lipoproteins was comparable with both plasma separation methods. Centrifugation was more time-efficient compared to filtration.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.