Abstract

ABSTRACTIt is important to analyze pyrolysis liquids to evaluate the yield of valuable products as well as unfavorable by-products. This work focuses on choosing detectors for reversed-phase ultra high-performance liquid chromatography analysis of pyrolysis liquids. The linearity, sensitivity, precision, and recovery of photodiode array (PDA) detector, single-quadrupole mass spectrometer (MS), and evaporative light scattering (ELS) detector were compared for the quantitative determination of several typical compounds found in pyrolysis liquids. PDA and MS detectors were found to be suitable for the quantification of furans and phenol derivatives (furfural, vanillin, syringol), but sugars and their derivatives (glucose, xylose, levoglucosan) can be analyzed with MS or ELS detectors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.