Abstract
ObjectivesThis study aimed to compare the performance of enzyme-linked immunosorbent assay (ELISA) and magnetic particle chemiluminescence immunoassay (MP-CLIA) for detecting anti-phospholipase A2 receptor (PLA2R) antibody and their clinical significance in idiopathic membranous nephropathy (IMN) patients. MethodsSerum samples from 448 patients with different types of nephropathy, including 222 with IMN, were tested using both methods. Sensitivity, specificity, and prognostic significance of PLA2R antibody levels were evaluated. ResultsSimilar sensitivity and specificity of ELISA and MP-CLIA in identifying IMN patients was found. However, MP-CLIA performed better than ELISA in predicting the prognosis of IMN patients. Adjusting the cutoff value reduced the false-negative rate in both methods. Logistic regression analysis identified six variables, including MP-CLIA-PLA2R levels, eGFR, lymphocyte count, B lymphocyte count, NK cell count, and complement 4, as predictors of renal function outcomes in IMN patients. ConclusionsBoth ELISA and MP-CLIA are reliable methods for detecting anti-PLA2R antibodies in IMN patients. However, MP-CLIA is more accurate in predicting the prognosis of IMN. Combining MP-CLIA-PLA2R with other variables can help predict renal function outcomes in IMN patients. Our study emphasizes the importance of considering both analytical performance and clinical utility when selecting a PLA2R antibody assay kit.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.