Abstract
Recent guidelines for infant cardiopulmonary resuscitation emphasize that all rescuers should minimize interruption of chest compressions, even for endotracheal intubation. We compared the utility of the Pentax-AWS Airway Scope (AWS) with an infant-sized Intlock (AWS-I), Airtraq laryngoscope (ATQ) and Miller laryngoscope during chest compressions on an infant manikin. Twenty-three novice doctors performed tracheal intubation on an infant manikin using the AWS-I, ATQ and Miller laryngoscope, with or without chest compressions. In Miller laryngoscope trials, one participant failed to secure the airway without chest compressions, while nine failed with compressions (P < 0.05). In ATQ trials, none of the participants failed without compressions, while six failed with compressions (P < 0.05). In AWS-I trials, all participants succeeded regardless of chest compressions. Intubation time was significantly longer with chest compressions with the Miller laryngoscope and ATQ, but not with the AWS-I. The AWS-I is an effective device for endotracheal intubation during chest compressions in infant simulations managed by novice doctors.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.