Abstract
The aim of this study is to compare effective dose (E) estimations based on different methods for patients with recurrent computed tomography (CT) examinations. Seventeen methods were used to determine the E of each phase as well as the total E of the CT examination. These included three groups of estimations: based on the use of published E, calculated from typical or patient-specific values of volume computed tomography dose index (CTDIvol) and dose-length product (DLP) multiplied by conversion coefficients, and based on patient-specific calculations with use of software.The E from a single phase of the examination varied with a ratio from 1.3 to 6.8 for small size patients, from 1.2 to 6.5 for normal size patients, and from 1.7 up to 18.1 for large size patients, depending on the calculation method used. The cumulative effective dose (CED) ratio per patient for the different size groups varied as follows: from 1.4 to 2.5 (small), from 1.7 to 4.3 (normal), and from 2.2 up to 6.3 (large). The minimum CED across patients varied from 38 up to 200 mSv, while the variation of maximum CED was from 122 up to 538 mSv.Although E is recommended for population estimations, it is sometimes needed and used for individual patients in clinical practice. Its value is highly dependent on the method applied. Individual estimations of E can vary up to 18.1 times and CED estimations can differ up to 6 times. The related large uncertainties should always be taken into account.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.