Abstract

The 1-bottle Oxoid Signal blood culture system was compared with a biphasic 2-bottle system in a clinical trial. A total of 851 routine blood samples were processed in the study which was run as a multicenter study at 4 university hospitals. Microorganisms were isolated from 104 (12.2%) blood cultures by one or both methods. 75 (72%) were clinically significant isolates, 61 of them were recovered in both systems, 9 in Oxoid Signal only and 5 in the biphasic system only. 29 isolates (28%) were judged as contaminants, 6 of them were found in both systems, 18 in Oxoid Signal only and 5 in the biphasic system only. The Oxoid Signal System had many advantages although it sometimes gave false positive signals in the device. The growth of pathogens was equally good, 93% versus 88% positive cultures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.