Abstract

The present study aimed to confirm the usefulness of a multi-laser handpiece system currently under development. Implants were placed in the tibia of rabbits using a conventional separate laser-implant handpiece system (control group; SurgicPro+; NSK, Kanuma, Japan and Epic 10; Biolase, Irvine, CA, USA) and a multi-laser handpiece system (experimental group; BLP 10; Saeshin, Daegu, Korea). Implants were placed in left and right tibias of five rabbits using a conventional laser-implant handpiece system and a multi-laser handpiece system (N = 5 per group). Subsequently, micro-computed tomography (micro-CT; bone-to-implant contact evaluation), implant stability quotient (ISQ) measurement, and histological evaluations were performed to confirm the implant placement results. The independent t-test and the paired t-test were used to compare the ISQ values and the results of the two implant-laser handpiece groups (α = 0.05), respectively. No statistically significant difference in micro-CT, ISQ, and histological evaluations was observed between implant placement by the two systems (p > 0.05) except implant initial stability. The use of the multi-laser handpiece system is expected to produce the same results as a conventional separate laser-implant handpiece system with the higher implant initial stability. Additionally, it will potentially make the clinical environment more pleasant and will provide convenience for the clinicians.