Abstract

To compare the objective visual quality of moderate-to-high myopia corrected by small incision lenticule extraction (SMILE) and transepithelial photorefractive keratectomy (TransPRK) at a 1,050-Hz ablation frequency, assisted by Smart-Pulse technology (SCHWIND eye-tech-solutions). This study involved 123 patients (123 eyes) with moderate-to-high myopia between July 2020 and January 2021. They were categorized into the SMILE group (67 patients, 67 eyes) and the TransPRK group (56 patients, 56 eyes). Follow-ups were conducted at 6 months postoperatively to record the logarithm of the minimum angle of resolution visual acuity, and the Strehl ratio and higher order aberrations were measured using the Sirius anterior segment analysis device (SCHWIND eye-tech-solutions) under a 6-mm pupil diameter at various postoperative intervals. At 1 week and 1 month postoperatively, the uncorrected distance visual acuity (UDVA) in the SMILE group was superior to that in the TransPRK group (P < .05 for both). At 1 week and 1 month postoperatively, the Strehl ratio value in the SMILE group was higher than that in the TransPRK group (P < .05 for both). At 1, 3, and 6 months postoperatively, coma was greater in the SMILE group than in the TransPRK group (P < .05 for all). Spherical aberrations were lower in the SMILE group than in the TransPRK group at 3 and 6 months postoperatively (P < .05). At 6 months postoperatively, UDVA was -0.09 ± 0.08 and -0.11 ± 0.05 logMAR in the SMILE and TransPRK groups, respectively, which exceeded their preoperative corrected distance visual acuity of -0.05 ± 0.04 and -0.09 ± 0.08 logMAR (all P < .001). Compared with preoperative values, the Strehl ratio, total higher order, coma, and spherical aberration differences were significantly increased postoperatively in both groups (all P < .001). Both surgical methods improved UDVA and each had its advantages. The visual quality of SMILE was superior at 1 week and 1 month postoperatively (Strehl ratio values were higher than those of the TransPRK group), and its spherical aberration was lower than that of the TransPRK group at 3 and 6 months; TransPRK with SmartPulse technology with a 1,050-Hz ablation frequency showed that coma was significantly lower than that of the SMILE group at 1, 3, and 6 months postoperatively. [J Refract Surg. 2024;40(7):e490-e498.].

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call