Abstract

We experimentally compared the air temperatures measured by non-contact imaging technique using parallel phase-shifting digital holography with those measured with a thermocouple. This study revealed that the technique has a faster thermal response than thermocouples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.